Log on to 7500 computer with your own user ID and password (will be created for you after you fill out the order form)
- To start the SDS (Sequence Detection Software), select:
Start>Programs>Applied Biosystems7300/7500>Applied Biosystems 7300/7500 SDS Software
- Select File>New.
- Select Plus/Minus in the Assay drop-down list of the New Document Wizard.
Accept the default settings for Container and Template (96-well Clear and Blank Document).
- Enter a name in the Default Plate Name field or accept the default.
- Click Next.
-
Select detectors to add to the plate document:
- click to select a detector.
- click Add. The detectors are added to the plate document.
- click Next.
-
Specify the detectors and tasks for each well:
- click on a well to select it.
- click on the detector name to select the detector for the well.
- click under the Task column to assing the detector task.
- Enter a quantity for wells that contain standards.
- click Use.
- click Finish.
-
Enter the sample names:
- select View>Well Inspector.
- click on a well or click-drag to select replicate wells.
- if necessary, change the settings for the Passive Reference dye, or ROX will be used.
- repeat steps b through d until you have specified sample names and passive reference dyes for all the wells on the plate.
- click (x) to close the Well Inspector.
- Verify the information on each well in the Setup tab.
- Click Pre-read to perform the pre-read run.
Next Topic: Amplification Run